Slide 8
Slide 8 text
BLA Measurement Improving on the art Take-away
The HERFD measurement
Optics This is an XAS scan, so we moved the monochromator
with coordinated motion of the undulator
Focusing We used the functional equivalent of a KB mirror pair,
which was a non-standard beamline component
Motors Sample stage + three degrees of freedom for the
spectrometer (linear rowards sample, crystal rotation,
rotation of Pilatus arm)
Sample Solution samples in plastic frames, enclosed with Kapton
tape
Scan file File containing columns of I0
, It
, ROI from the SDD, some
areal ROIs from camserver, & a few miscellaneous
columns. Also a column of metadata!
Images One image for each energy pointin the XANES scan +
10-15 images for energy calibration
8 / 20
Spectrometry, Data Acquisition, and 2D Image Processing